3Ć51

Maintaining the System

Cleaning the Roller Cages

Use this procedure to clean the roller cages for the reagent, sample, or waste
pumps.

Materials required:

S 10% solution of household bleach

NOTE:

Dilute household bleach (5.25% sodium hypochlorite) 1:10 with

reagent-quality water.

S reagent water
S lint-free tissue and swabs

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Stop the system from the Menu screen:

a. Select

2 Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

2.

Disconnect the pump tubing from the reagent manifold.

NOTE:

Do not stretch the tubing.

3.

While holding the left side of the tubing, turn the roller cage clockwise and
gently pull the tubing away from the platen and the roller cage.

4.

Set the tubing assembly aside.

5.

Grasp the roller cage and gently pull it straight off its shaft as shown in
Figure 3-33.

2

Menu Code

7