3Ć87

Maintaining the System

CAUTION:ĄĂ

Do not remove or return the sensors to the measurement module

without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.

2.

Remove the biosensors as shown in Figure 3-60:

Figure 3Ć60.āăRemoving the Glucose and Lactate Biosensors

Grasp the tab on the sensor and pull the sensor

up and out of the measurement module.

Push the springĆloaded latch to the

right.

a. Push up the latches on the measurement module door and lift the door.

b. Push the spring-loaded latch to the right.

c. Grasp the tab on the biosensor and pull the biosensor up and out of the

measurement module.

d. Discard the biosensor.

3.

Remove the new biosensor from the foil package.

4.

Ensure that the O-ring is in place on the biosensor.

CAUTION:

Do not remove or return the sensors to the measurement module

without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.

NOTE:

As you install the biosensors, ensure that the biosensors are in the correct

location. Visually verify that you align the contacts on the biosensors with the
contacts in the measurement module. Slide the remaining sensors to the right to
create more space, if needed.

5.

Install the biosensor:

!

!