Bayer HealthCare Rapidlab 800 User Manual

Page 148

Advertising
background image

2Ć50

Rapidlab 800 Operator's Manual

This flag . . .

Indicates . . .

the result is below the lower limit of the target range.

↑↑

the result is above the upper limit of the action range.

↓↓

the result is below the lower limit of the action range.

- - -

the result is above the upper limit of the measurement range.

- - -

the result is below the lower limit of the measurement range.

The following table describes the messages that may appear in QC sample reports
on the roll printer:

Message

Description

****** = Not in Calibration

Sensor is out of calibration.

* = D5 No Endpoint: __

QC sample analysis did not reach endpoint.

or

= exceeds target range

QC sample result is above or below the target range.

↑↑

or

↓↓

= exceeds action

range

QC sample result is above or below the action range.

ACCEPTED INTO QC FILE
__

Indicates the QC result status (accepted) and the file
number in which the results are stored.

Bubbles Detected In Sample

System detects a non-continuous fluid in the
measurement module sample path.

D2 Excessive Drift: __

Sensor drift is beyond predefined limits during a
one-point or a two-point calibration.

D38 Temperature Error

System detects an error in the temperature control
system.

DISCARDED INTO QC
FILE __

Indicates the QC result status (discarded) and the file
number in which the results are stored.

REJECTED INTO QC
FILE __

Indicates the QC result status (rejected) and the file
number in which the results are stored.