3Ć99

Maintaining the System

Replacing the Sample Probe

Materials required:

S sample probe

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Stop the system from the Menu screen:

a. Select

2 Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

2.

Remove the old probe:

a. Disconnect the sample tubing from the connector on the measurement

module.

b. Disconnect the sample tubing from the bracket on the probe mount.

c. Rotate the sample tubing toward you by pulling it up until it is

perpendicular to the bench top as shown in Figure 3-67.

Figure 3Ć67.āăRotating the Sample Tubing

Sample Tubing

Sample Probe

Bracket

Probe Mount

Sample Tubing Connector

d. Disconnect the sample tubing from the sample probe.

e. Pull the probe tab to the left into the slot, and pull the probe out of the

sample port, as shown in Figure 3-68.

2

Menu Code

7