3Ć79

Maintaining the System

Figure 3Ć53ā.ăFilling the Electrode Compartment

Electrode

Compartment

OĆring

Sample Path

4M KCl Fill Solution

OĆring

Reference Sensor Cassette

(KCI Reservoir)

CAUTION:

Do not touch the internal electrode wire. The wire is fragile and is

easily damaged.

8.

Use the hex tool to remove the new internal electrode from its container.

9.

Insert the internal electrode into the compartment and screw it into place with
the hex tool, ensuring that you do not cross-thread the electrode.

10. Fill the KCl reservoir in the sensor:

a. Remove the reservoir cap with the hex tool and set it aside as shown in

Figure 3-54.

Figure 3Ć54.āăRemoving the Reservoir Cap with the Hex Tool

Hex Tool

Reservoir Cap

OĆring

OĆring

Sample Path

KCI Reservoir

Vent Hole

b. Partially fill the KCl reservoir by gently squeezing the KCl fill solution

container as shown in Figure 3-55.

!