Troubleshooting the System

Valve

Function

Vent

controls the flow of ambient air into the manifold

Clean

controls the flow of Cleaning Solution

G/L

controls the flow of Cal G/L Reagent

Bypass

diverts the Wash G/L Zero Reagent in the reagent manifold when the
foam wash valve is off, preventing a pressure buildup

Diverter

directs the flow of reagent into either the calibration passage or the
wash passage of the sample port

If you want to test the Cal Gas or Slope Gas valves, refer to Checking the Gas
Flow
, page 4-64.

1.

Access the Valves Test from the Menu screen:

a. Select

3 Troubleshooting

and press

Enter

.

b. Select

1 Fluidics System

and press

Enter

.

c. Select

4 Valves

and press

Enter

.

The Valve Test screen appears as shown in Figure 4-6.

Figure 4Ć6.ā Valve Test Screen on an 860

Test one valve

at a time.

Check here for results.

2.

Select the valve you want to test and press

Enter

.

3.

Press

Start Test

.

Listen for a distinctive click, which indicates the opening of the valve, and
check the screen for the message, Valve energized, to verify that the valve is
operating correctly.

4.

Press

Stop Test

.

Listen for a distinctive click, which indicates the closing of the valve.