System messages – Bayer HealthCare Rapidlab 800 User Manual

Page 288

Advertising
background image

4Ć6

800 Series Operator's Manual

System Messages

System messages provide information about the operating status of the system and
can appear in the status area of the screen, in the Status Event Log, or on printed
reports. Table 4-1 describes each system message.

Table 4Ć1.āăSystem Messages

System Message

Description

1-pt Cal Pending
in __ Min

Appears 5 minutes before a one-point calibration is due and
counts down until the calibration starts.

1-pt Metabolite Cal
Due

Appears during an analysis to indicate the system will perform a
metabolite calibration at the completion of the analysis.

1-pt Metabolite Cal
Due in __ Min

Appears 5 minutes before a one-point metabolite calibration is
due and counts down until the calibration starts.

2-pt Cal Pending
in __ Min

Appears 5 minutes before a two-point calibration is due and
counts down until the calibration starts.

Additional
Messages

Appears in the status area to indicate that there are more
messages, which can be viewed in the status log.

Auto Clean
Pending in __ Min

Appears 5 minutes before an Auto Clean sequence is due and
counts down until the sequence starts.

Bubbles Detected
In Sample

Appears when the system detects a non-continuous fluid in the
measurement module sample path. See Troubleshooting System
Messages
, page 4-114.

Calibration
Overdue: __

Appears when the system must perform a calibration on the
sensor(s) indicated.

COox Cover Open
During Meas

Appears when the cover of the CO-ox module is open during a
sample or slope measurement. Question the results.

COox Cover Open
During Zero

Appears when the CO-ox module cover is open during the CO-ox
module zero measurement sequence of a one- or two-point
calibration. Repeat the calibration with the CO-ox cover closed.

COox Cover is
Open

Appears when the cover of the CO-ox module is open when you
begin the sample measurement. You must close the cover before
pressing Analyze to ensure an accurate CO-ox analysis.

(Continued)