Precision on controls – Bayer HealthCare Rapidlab 800 User Manual

Page 574

Advertising
background image

EĆ48

800 Series Operator's Manual

Substance

Level of Interference

Interferes with . . .*

Fetal Hemoglobin at 80%

tHb

COHb

MetHb

HHb

–0.9 g/dL

+3.1%

–1.1%

–2.6%

Carboxymethylcellulose

tHB

COHb

O

2

Hb

HHb

–2.0 g/dL

+10.0%

–10.0%

+6.5%

* Analytes that are within interference criteria stated in Table EĆ40 are not listed.

Precision on Controls

Quality control materials were analyzed on the 800 series CO-ox module. The
results are presented here.

Precision on aqueous quality control materials was estimated using four 800
systems. One run per instrument was made over a eight day period. Three replicates
of each control level were analyzed in each run.

Table E-43 summarizes the results of the 800 CO-ox module precision on QC
materials.

Table EĆ43.āă800 COĆox Module Precision Results

Parameter

Level

n

Mean

WRSD

*

TotSD

[

tHb

1

96

9.36

0.05

0.25

2

96

13.98

0.07

0.37

3

96

21.60

0.08

0.54

FO

2

Hb

1

96

5.79

0.11

0.28

2

96

13.56

0.18

0.30

3

96

22.01

0.22

0.30

(Continued)