Recovery and precision with whole blood – Bayer HealthCare Rapidlab 800 User Manual

Page 575

Advertising
background image

Performance Characteristics

EĆ49

Parameter

TotSD

[

WRSD

*

Mean

n

Level

FCOHb

1

96

52.64

0.18

0.35

2

96

32.03

0.16

0.30

3

96

9.52

0.35

0.39

FMetHb

1

96

5.05

0.13

0.23

2

96

4.70

0.14

0.24

3

96

4.07

0.19

0.30

FHHb

1

96

36.52

0.30

0.44

2

96

49.71

0.25

0.36

3

96

64.40

0.30

0.41

* WRSD = within-run standard deviation

[

TotSD = total standard deviation

Recovery and Precision with Whole Blood

For testing syringe and capillary modes, blood was collected in heparinized vacuum
tubes. It was tonometered at 37

°

C and/or adjusted chemically. Three levels of tHb

were prepared by separating the red cells from the plasma and recombining
appropriately. For each run, three replicates of each level were run on four 800
systems.

Table E-44 through Table E-46 summarize the results of the 800 CO-ox module
recovery and precision testing.

Table EĆ44.āă800 COĆox Module Recovery and Precision with Total Hemoglobin of

5-8 g/dL and Oxyhemoglobin > 80%

Analyte

Mode

n

Mean

Reference

Mean Difference

from Reference

WRSD

*

tHb

Capillary

72

6.64

–0.19

0.30

Syringe

66

6.53

–0.27

0.11

FO

2

Hb

Capillary

72

91.96

1.16

0.45

(Continued)

Advertising