Bayer HealthCare Rapidlab 800 User Manual

Page 114

Advertising
background image

2Ć16

Rapidlab 800 Operator's Manual

Figure 2Ć10.ā Sample Type Menu

Select pH Only.

3.

Select

pH Only

and press

Enter

.

4.

Press

Analyze

.

NOTE:ĄĂ

Ensure that the trailing edge of the sample remains in contact with the

sample ground/temperature sensor.

5.

Turn the sample pump clockwise until the leading edge of the sample fills the
reference sensor, as shown in Figure 2-11.

Figure 2Ć11.ā Sample at Reference Sensor on an 850

Reference Sensor

Sample Ground/
Temperature Sensor

Sample Pump

6.

Press

Sample in Place

.