2Ć22

Rapidlab 800 Operator's Manual

Analyzing Samples to Combine with

270 COĆoximeter Results

Use this procedure to analyze a patient sample at an 840, 850, or 860 system and
combine the results with results from a sample analyzed at a 270 CO-oximeter that
is connected to one of these systems. Refer to Appendix D, Connecting to External
Devices
, for information about connecting the 270 CO-oximeter to an 800 base
model.

To combine results, you must simultaneously analyze the sample at the 800 system
and the 270 CO-oximeter. The 800 system assigns a sequence number to each
sample and reports sample results by this number. The 800 system stores the 270
results and prints a report showing the blood gas, CO-oximeter, and combined
results. The 800 also stores any pass-through results from the 270. You can recall
these results from the 800 system. Refer to Appendix F, Printed Reports, to see an
example of a combined report.

If you want the system to analyze a different set of parameters for the sample,
press Change Parameters to select another panel. Refer to Selecting Parameter
Panels
, page 2-29, for more information.

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Prepare the sample and introduce the sample at the 270 CO-oximeter.

2.

Insert the device into the sample port at the 840, 850, or 860 system.

3.

Press

Analyze

.

4.

When prompted, remove the sample device.

5.

When prompted, remove the sample device.

CAUTION:ĄĂ

If your laboratory requires you to enter an ID number at each

system, be sure that you enter the same ID number at each system. The patient
ID you enter at your 800 system and the sample ID you enter at your 270
CO-oximeter must match to combine results.

6.

Return to the 800 system and type the required information in the Patient
Information screen. Press

Done

when you finish.

Refer to Entering Patient Sample Data, page 2-27, if you need more
information to complete this screen.

7.

If required by your laboratory, type the sample ID at your 270 CO-oximeter.

When the CO-oximeter data is available, the 800 system displays the final
results and prints the report.

8.

Press

Home

to return to the Ready screen. If you do not press Home, the

system automatically returns to the Ready screen after 45 seconds.

!