1Ć17

Learning About the System

Figure 1Ć9.ā 860 Measurement Module Components

Lactate Sensor

pH Sensor

Reference Sensor

Spring-Loaded Latch

Measurement
Module Tubing

Preheater

pO

2

Sensor

pCO

2

Sensor

Sample Ground/
Temperature Sensor

Glucose Sensor

K+ Sensor

Ca++ Sensor

Cl – Sensor

Na+ Sensor

Release Tab

Contact Assembly

Measurement
Block

COĆox Module Measurement Components

The CO-ox module is connected to the base model, which supplies power to the
module. The 844, 854, and 864 systems analyze samples for tHb and FO

2

Hb; the

845, 855, and 865 systems anlayze samples for tHb, FO

2

Hb, FHHb, FMetHb, and

FCOHb. The measurement components spectrophotometrically measure tHb and
its derivatives. Most of the measurement components are enclosed in the CO-ox
module. The sample chamber is visible when the CO-ox cover is raised.

Optics Module and Lamp

The lamp provides the light to illuminate the sample. The optics module includes
lenses and filters, fiber optic coupler, and polychromator. The lenses and filters
focus the light from the lamp to the fiber optic coupler. The fiber optic coupler
transports the light to the sample chamber. The polychromator, which consists of
coupling lenses, entrance slit, mirrors, and diode array, disperses the light that
passes through the sample chamber into a spectrum and measures the light
intensity at several wavelengths.