5Ć28

NOTE:ĄĂ

The first operator password has a default value of 12345. To ensure

that there is always at least one active operator password, you can edit the
default password, but you cannot delete it.

Figure 5Ć15.ā Operator Passwords Screen

NOTE:ĄĂ

If you want to define only one operator password for all operators to

use, enter the password you want and leave the Operator ID field blank. The
operator ID in the Patient and QC Data Entry forms and the Maintenance Task
Done dialog box will be blank. The operator ID on the patient sample, QC
sample, and maintenance log reports will also be blank.

3.

Take the appropriate action.

If you want to . . .

Then . . .

define a new operator
password

a. Type the ID in the Operator ID field and press

Enter

.

b. Type the password in the Operator Password field

and press

Enter

.

edit a password or ID

retype the ID in the Operator ID field or the password
in the Operator Password field and press

Enter

.

delete a password or ID

press

Clear Entry

to delete text in the Operator ID or

Operator Password field.

Bayer Diagnostics recommends that you keep a record of operator IDs and
passwords to help you track and manage assigned operator passwords.
Operator Password Screens are numbered 1 to 50 and operator fields are
numbered 1 to 10 on each screen to help you list the passwords.

4.

Press

Done

.

5.

Define another setup function or press

Exit Menu

to return to the Ready

screen.