Replacing the capillary seal – Bayer HealthCare Rapidlab 800 User Manual

Page 272

Advertising
background image

3Ć100

Rapidlab 800 Operator's Manual

Figure 3Ć68.āăReplacing the Sample Probe

Sample Probe

Probe Tab

Slot

Bracket

Sample Tubing Connector

3.

Install the new probe:

CAUTION:ĄĂ

Insert the probe slowly to avoid bending the shaft.

a. Push the sample probe into the opening in the sample port and place the

probe tab in the slot.

b. Connect the sample tubing to the new sample probe.

c. Rotate the sample probe and tubing down toward the system.

d. Place the sample tubing in the bracket on the probe mount.

e. Connect the sample tubing to the sample tubing connector on the

measurement module.

4.

Press

Continue

.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

5.

Press

Yes

to perform a two-point calibration.

Replacing the Capillary Seal

Materials required:

S capillary seal

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Stop the system from the Menu screen:

!

2

Menu Code

7