Pump functions test – Bayer HealthCare Rapidlab 800 User Manual

Page 344

Advertising
background image

4Ć58

800 Series Operator's Manual

3.

Check the screen for the message, Reagent pump flow rate acceptable, to
verify that the flow rate is within the acceptable range.

If the flow rate is . . .

Then . . .

acceptable

press

Exit Test

.

unacceptable

a. Press

Calibrate Pump

.

b. Check the screen for the message, Pump calibration

complete.

c. Press

Exit Test

.

4.

Press

Exit Menu

.

Pump Functions Test

Use this procedure to test the ability of the reagent, sample, and waste pumps to
turn on and off and to manually test the flow rates for the reagent, sample, waste,
and CO-ox pumps. The flow rate is the amount of fluid the pumps can pump in a
specified time.

Materials required for the manual pump flow rate test:

S stopwatch
S 10 mL graduated cylinder
S small container of reagent water
S 2 small two-way connectors
S 2 pieces of test tubing, each approximately 30 cm (10 inches) long with

approximately 0.15 cm (0.060 inches) inner diameter

1.

Access the Pump Functions Test from the Menu screen:

a. Select

3 Troubleshooting

and press

Enter

.

b. Select

1 Fluidics System

and press

Enter

.

c. Select

3 Pump Functions

and press

Enter

.

The Pump Functions Test screen appears as shown in Figure 4-4.

3

1

Menu Code

3