EĆ44

800 Series Operator's Manual

Table EĆ38.āă860 Recovery and Precision TestingĊGlucose

*

Level

Mode

n

WRSD

[

Observed

Expected

Recovery

45

Syringe

39

0.9

47

46

102.8

Capillary

30

1.4

46

43

107.9

Microcapillary

30

1.8

44

43

101.2

Microsyringe

36

1.2

45

45

100.6

pH/Lyte

21

1.6

47

47

100.7

81

Syringe

33

2.1

85

84

100.6

Capillary

33

1.4

85

83

102.5

Microcapillary

36

2.5

79

79

99.8

Microsyringe

45

1.9

84

82

102.3

pH/Lyte

24

1.1

81

80

101.1

207

Syringe

33

3.5

205

205

100.4

Capillary

33

2.9

206

203

101.5

Microcapillary

36

3.9

206

205

100.8

Microsyringe

45

3.8

205

203

101.2

pH/Lyte

23

3.1

206

205

100.6

* Performance determined with metabolite recal on. Refer to Glucose Biosensor Calibration in Section 1

for information on performance with recal off.

[

WRSD = within-run standard deviation