Bayer HealthCare Rapidlab 800 User Manual

Page 228

Advertising
background image

3Ć58

Rapidlab 800 Operator's Manual

3.

Deproteinize the sample path:

a. Prepare the deproteinizer as directed on the package.

b. Select

2 Maintenance

and press

Enter

.

c. Select

1 Deproteinize

and press

Enter

.

d. Invert the deproteinizer vial several times to mix.

e. Insert an aspiration adapter into the sample port and insert the other end

into the deproteinizer.

f. Press

Analyze

.

g. When prompted, remove the adapter.

h. Wait for the deproteinizing cycle to finish.

A message box appears prompting you to condition the sensors.

4.

Perform the cleaning cycle:

a. Press

Yes

.

b. Insert an aspiration adapter into the sample port and immerse the other end

in the glutaraldehyde solution.

c. Press

Analyze

.

d. When prompted, remove the adapter.

e. Wait 5 minutes for the cleaning cycle to finish.

NOTE:ĄĂ

Press

Cancel

if you want to stop cleaning.

The cleaning cycle finishes. The system performs an extended wash sequence.
The Calibrate System message box appears at the end of the wash sequence.

5.

Press

No

.

6.

Initiate a wash sequence from the Menu screen:

a. Select

2 Maintenance

and press

Enter

.

b. Select

8 Wash

and press

Enter

.

NOTE:ĄĂ

As you install the biosensors, ensure that the biosensors are in the

correct location. Visually verify that you align the contacts on the biosensors
with the contacts in the measurement module.

CAUTION:ĄĂ

Do not remove or return the sensors to the measurement module

without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.

7.

When the wash sequence finishes, remove the test/blank glucose and lactate
sensors and reinstall the biosensors as described in

Replacing the Glucose and

Lactate Biosensors

page 3-86.

2

Menu Code

1

2

Menu Code

8

(from the Main Menu)

!