Rapidlab 800 Operator's Manual

Daily Maintenance

The daily maintenance schedule is based on analyzing 30 samples per day, unless
otherwise noted. If your laboratory analyzes more than 30 samples per day,
perform daily maintenance more frequently.

Cleaning the Exterior Surfaces

Materials required:

S 10% solution of household bleach

NOTE:

Dilute household bleach (5.25% sodium hypochlorite) 1:10 with reagent

quality water.

S reagent-quality water

44

S lint-free tissues or swabs

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

S Wipe with a 10% solution of household bleach all exterior surfaces, including:
S the sample entry components and the drip tray
S the waste area