Bayer HealthCare Rapidlab 800 User Manual

Page 485

Advertising
background image

System Administration

5Ć75

4.

Complete the restore process.

If . . .

Then . . .

a message appears
prompting you to
insert the next diskette

a. Remove the diskette from the diskette drive.

b. Insert another backup diskette and press

Continue

.

the Restore Finished
screen appears

Continue with step 5.

5.

Press

OK

and the remove the diskette.

The system performs an automatic shutdown.

The Shutdown screen appears.

6.

Press

Yes

to shut down the system.

CAUTION:ĄĂ

You must wait for at least 1 minute before you disconnect the

power cord and then wait at least 10 seconds before you reconnect the power
cord. If you do not adhere to the time intervals, you can damage the system.

A message appears on the screen and on the roll printer, directing you to wait
before you disconnect the power.

7.

Wait at least 1 minute.

8.

Disconnect the power cord from the power supply.

9.

Restart the system:

a. Wait at least 10 seconds after disconnecting the power cord.

b. Reconnect the power cord to the power source.

After a few moments, the system starts initializing. When initializing is
complete, a screen for Analyze Mode appears.

Ensure that you restore data to the system hard disk only from the backup
diskettes that are labeled for that system.

If you insert a diskette that is not a backup diskette, the Invalid Diskette Data
type message appears.

If . . .

Then . . .

you want to continue

insert the correct diskette and press

Continue

.

you do not want to
continue

Press

Cancel

.

The Cancel Restore message appears.

!

Procedural

Notes