2Ć15

Operating the System

Analyzing pH Only Samples

Since pH samples are frequently capillary tubes with insufficient sample volume,
the system requires you to move the sample manually to the measurement module.
If the sample device is a syringe that contains sufficient volume for analysis, the
system positions the sample automatically. The following table lists the minimum
sample volumes required by each system to determine pH:

Table 2Ć4.āăMinimum Sample Volumes

System

Minimum Sample Volume

840, 844, 845

35

µ

L

850, 854, 855

70

µ

L

860, 864, 865

95

µ

L

NOTE:

When you select pH only, the system measures everything except pCO

2

and pO

2.

On systems with the CO-ox module, the system measures everything

except pCO

2

, pO

2

, and tHb parameters.

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

CAUTION:

Insert the fire-polished end of the capillary tube into the sample port to

prevent damage to the capillary seal.

1.

Prepare the sample and insert the device in the sample port. Figure 2-9 shows
a capillary sample in the sample port.

Figure 2Ć9.ā Inserting a Capillary Sample

Sample Port

Capillary Sample

2.

Press

Change Sample Type

.

The Sample Type menu appears, as shown in Figure 2-10.

860

850

!