Cleaning the hemolyzer – Bayer HealthCare Rapidlab 800 User Manual

Page 198

Advertising
background image

3Ć30

Rapidlab 800 Operator's Manual

Cleaning the Hemolyzer

Materials required:

S lint-free tissue or lint-free swabs
S reagent-quality water
S dilute cleaning solution (do not use bleach, alcohol, or abrasive powder)
S hemolyzer gasket

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

1.

Stop the system from the Menu screen:

a. Select

2

Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

2.

Disassemble the hemolyzer as shown in Figure 3-16:

Figure 3Ć16.āăDisassembling the Hemolyzer

Anvil Cap

Anvil

Sample Tubing

Anvil Spring

Mounting Pin

Hemolyzer

Gasket (Anvil Seal)

a. Remove the anvil cap on the hemolyzer by turning it one quarter-turn

counterclockwise.

b. Disconnect the sample tubing from the anvil.

c. Pull the anvil and anvil spring away from the mounting pin on the

hemolyzer.

2

Menu Code