D22 barometric pressure error, Problem, Solutions – Bayer HealthCare Rapidlab 800 User Manual

Page 318

Advertising
background image

4Ć34

800 Series Operator's Manual

D22 Barometric Pressure Error

Problem

The barometer detects atmospheric pressure beyond predefined limits.

Solutions

Check that the system barometer is functioning properly:

1.

From the Menu screen, select

3 Troubleshooting

and press

Enter

.

2.

Select

2 Temp/pAtm

and press

Enter

.

3.

Press

Start Test

.

4.

Check the screen to verify whether the atmospheric pressure (pAtm) is
within the range of 400 to 825 mm Hg (53 to 110 kPa) and that the reading
is stable.

5.

Compare the atmospheric pressure displayed to the barometer in your
laboratory to verify its accuracy.

6.

Press

Stop Test

.

7.

Press

Exit Test

.

If the reading is stable but does not match the barometer in your laboratory,
calibrate the barometer:

8.

From the Menu screen, select

1 Calibration

and press

Enter

.

9.

Select

8 Barometer

and press

Enter

.

10. Type the correct atmospheric pressure and press

Enter

.

11. Press

Done

.

12. Perform a two-point calibration.

If the D code reappears, contact your Service Representative.

D22

3

Menu Code

2

1

Menu Code

8