Performing a barometer calibration – Bayer HealthCare Rapidlab 800 User Manual

Page 157

Advertising
background image

2Ć59

Operating the System

Performing a Barometer Calibration

Use this procedure to calibrate the internal atmospheric pressure sensor to a
barometer in your laboratory.

1.

Access the Barometer Calibration screen from the Menu screen:

a. Select

1 Calibration

and press

Enter

.

b. Select

8 Barometer

and press

Enter

.

2.

Type the correct atmospheric pressure and press

Enter

.

The acceptable range for atmospheric pressure is 400 – 825 mm Hg

(53.0 – 110.0 k Pa).

3.

Press

Done

to save the new atmospheric pressure and return to the Ready

screen.