Bayer HealthCare Rapidlab 800 User Manual

Page 161

Advertising
background image

2Ć63

Operating the System

Figure 2Ć35.ā Calibration Data Search Result Screen

This line contains

the calibration type,

sequence number,

and date and time

of the calibration.

Press Reporting Options to

print or transmit this report.

6.

Perform the appropriate task at the Calibration Data Search Result screen.

If you want to . . .

Then . . .

print the calibration report

a. Press

Reporting Options

.

b. Press

OK

.

The calibration report prints, as

shown in Figure 2-36.

transmit the results to an LIS
or a data management
system

a. Press

Reporting Options

.

b. Select Send Results and press

Enter

.

c. Press

OK

.