D19 Fluid Detector Error:

Problem

Fluid detector millivolt (mV) reading is beyond predefined limits.

Qualifier:

FD5

Solutions

Check the CO-ox sample path for obstructions. If you see any obstructions,
remove them as described in Removing Obstructions from the CO-ox Sample
Path
, page 4-87.

Check the sample tubing that goes through FD5 for cracks, leaks, and
discoloration. If you see cracks or leaks, replace the tubing as described in
Replacing the CO-ox Sample Tubing in Section 3.

Perform the Fluid Detector Test as described in Fluid Detector Test, page 4-66.

Deproteinize the sample path as described in Deproteinizing the Sample Path in
Section 3 and repeat the Fluid Detector Test.

If the D code reappears, contact your Service Representative.

D19