Performance Characteristics

EĆ19

Parameter

TotSD

[

WRSD

*

Mean

n

Level

pO

2

1

44

25.8

1.39

1.97

4

44

237.7

2.99

5.78

Na

+

1

44

102.4

0.34

1.31

4

44

167.2

0.48

1.43

K

+

1

44

2.04

0.059

0.073

4

44

16.87

0.170

0.235

Cl

1

44

132.1

0.95

1.22

4

44

75.9

0.47

0.52

Ca

++

1

44

2.82

0.035

0.056

4

44

0.55

0.009

0.017

* WRSD = within-run standard deviation

[

TotSD = total standard deviation

Recovery and Precision with Whole Blood and Expired

Gases

For testing syringe, capillary, microsyringe, and microcapillary modes, blood was
collected in heparinized vacuum tubes. It was tonometered at 37.0

_C to each of

three levels to prepare samples for pH analysis, and five levels to prepare samples
for pCO

2

and pO

2

analysis. Blood was spiked/diluted to each of three levels to

prepare samples for sodium, potassium, calcium, and chloride analysis. Multiple
runs were made using these samples on four 850 systems. The experimental
protocol called for three replicates of each level in each run.

For testing the expired gas mode, 10 mL of tonometry gas were drawn into a
12 mL syringe and aspirated into the 850 system for analysis. Multiple runs were
made using five levels of expired gas on four 850 instruments. The experimental
protocol called for three replicates of each level in each run.