Bayer HealthCare Rapidlab 800 User Manual

Page 248

Advertising
background image

3Ć76

Rapidlab 800 Operator's Manual

Figure 3Ć49.āăRemoving the Reservoir Cap with the Hex Tool

Hex Tool

Reservoir Cap

OĆring

OĆring

Sample Path

KCI Reservoir

Vent Hole

b. Partially fill the KCl reservoir as shown in Figure 3-50.

Figure 3Ć50.āăFilling the KCl Reservoir

4M KCl Fill Solution

Fill Line

OĆring

OĆring

Sample Path

Electrode

Compartment

c. Tap the front face of the sensor with your knuckle to release any bubbles.

d. Gradually fill the KCl reservoir with KCl fill solution up to the fill line.

CAUTION:ĄĂ

Do not overtighten the reservoir cap. Overtightening can deform

the gasket and cause leaks.

e. Reinstall the reservoir cap and hand tighten.

10. Continue with

Reinstalling a Sensor

, page 3-89.

!