Controlling capillary sample movement – Bayer HealthCare Rapidlab 800 User Manual

Page 447

Advertising
background image

System Administration

5Ć37

Controlling Capillary Sample Movement

Use this procedure to control the way the system moves capillary samples to the
measurement module. You can choose to move the samples manually or to have
the system move them automatically.

Table 5Ć11.āăAuto Move Capillary Options

Option

Description

Yes

The system moves capillary samples automatically to the measurement
module. The default value is Yes.

No

The operator moves capillary samples to the measurement module by
turning the sample pump.

1.

Access the System Options screen from the Menu screen:

a. Press

6 System Setup

and press

Enter

.

b. Press

5 Systems Options

and press

Enter

.

The System Options screen appears, as shown in Figure 5-22.

Figure 5Ć22.ā System Options Screen for an 850

Move to the Auto Move

Capillary frame and

select Yes or No.

2.

Move to the Auto Move Capillary frame.

3.

Select

Yes

or

No

and press

Enter

.

4.

Press

Done

.

5.

Define another setup function or press

Exit Menu

to return to the Ready

screen.

6

Menu Code

5