1.

Access the System Information screen from the Menu screen:

a. Press

8 Service Setup

and press

Enter

.

b. Press

1 System Information

and press

Enter

.

The System Information screen appears, as shown in Figure 5-1.

Figure 5Ć1.ā System Information Screen for an 850

Total cycle count is the

number of patient

samples, QC samples,

calibrations, and

operatorĆinitiated

washes.

2.

Press

Menu

when you finish viewing the screen.

3.

Define another setup function or press

Exit Menu

to return to the Ready screen.

Defining the Patient Information Screen

Use this procedure to select the fields you want to use in the Patient Information
screen. The Patient Information screen can contain 16 fields. The default for each
field is On, which means that all fields appear on the Patient Information screen.
You can choose to have a field not appear on the form by turning the field off.

Certain fields can be made required fields. Required fields are fields that must be
completed to continue with analysis. You can make a field required by selecting the
check box next to the field. The default for required fields is Off, which means that
none of the fields are required. Table 5-2 describes the Patient Information fields.

Table 5Ć2.āăPatient Information Fields

Field

Description

Patient ID

number used to identify the patient

Accession #

number used to identify the sample

Location

location of the patient

(Continued)

8

Menu Code

1