D70 optics error, Problem, Solutions – Bayer HealthCare Rapidlab 800 User Manual

Page 328

Advertising
background image

4Ć44

800 Series Operator's Manual

D70 Optics Error:

Problem

The system detects an error in the CO-ox optical measurement system.

Qualifiers: 2

3

4 7

9

Solutions

Qualifier

Procedures

2

Perform a pH/Lytes one-point calibration.

Replace the 7.3 Buffer/CO-ox zero reagent.

Check that the area around the sample chamber is free from lint or dust.

Prime and perform a pH/Lytes one-point calibration.

Clean the sample chamber as described in Cleaning the Sample
Chamber.
It is not necessary to remove or replace the sample chamber
gasket.

3

Perform a pH/Lytes one-point calibration.

Perform CO-ox Optics test.

Clean the sample chamber as described in Cleaning the Sample
Chamber.
It is not necessary to remove or replace the sample chamber
gasket.

4

Ensure the CO-ox cover is closed.

Perform a pH/Lytes one-point calibration.

7

Ensure the CO-ox cover is closed.

9

Contact your Service Representative.

If the D code reappears, contact your Service Representative.

D70