Performing thb slope calibrations – Bayer HealthCare Rapidlab 800 User Manual

Page 155

Advertising
background image

2Ć57

Operating the System

Performing tHb Slope Calibrations

Use this procedure to calibrate the tHb slope for the CO-ox module. You can
perform this calibration from either the Analyze mode or
the Menu mode.

1.

Initiate the tHb slope calibration procedure.

If you are in the . . .

Then . . .

Analyze mode

a. Press

Calibrate

.

b. Select

tHb Slope.

Menu mode

a. Select

1 Calibration

and press

Enter

.

b. Select

9 tHb Slope

and press

Enter

.

The tHb Slope Calibration screen appears, as shown in Figure 2-33.

Figure 2Ć33.ā tHb Slope Calibration Screen

2.

Enter the slope value for the slope reagent and press

Enter

.

NOTE:ĄĂ

The last-entered value is displayed. To remove this value, press the

Clear Entry

F-key and enter the new value.

3.

Press

Done

.

4.

Insert the sample device into the sample port.

5.

Press

Analyze

.

6.

When prompted, remove the sample device.

The system performs a wash at the end of the calibration and then returns to
the Ready screen.