Replacing 7.3/coćox zero and cal g/l reagents – Bayer HealthCare Rapidlab 800 User Manual

Page 188

Advertising
background image

3Ć20

Rapidlab 800 Operator's Manual

NOTE:

Do not remove or tighten the cap that contains the reagent septum.

Removing or tightening the cap damages the integrity of the reagent septum.

4.

Remove the plug from the cap of the new cleaning solution bottle.

5.

Insert the cleaning solution bottle into position on the reagent manifold.

6.

Push the bottle to ensure that it fits tightly on the reagent fitting.

7.

Press

Reset Levels

.

The Reset Levels screen appears.

8.

Select C1/C2 and press

Done

.

Replacing 7.3/COĆox Zero and Cal G/L Reagents

Use this procedure to discard and replace the 7.3/CO-ox Zero and Cal G/L reagents
if they are opened and installed on the system for 30 days.

Materials required:

S 7.3/CO-ox Zero (7.382 Buffer)
S Cal G/L

WARNING

Wear safety glasses, gloves, and a laboratory coat when handling the

reagents.

1.

At the Ready Screen, press

Enter

.

The Reagent Levels screen appears.

2.

Remove the reagent bottles from the reagent manifold as shown in Figure 3-8
for an 860 system.

Figure 3Ć8ā.ăRemoving the 7.3/COĆox Zero and Cal G/L Reagents

Cal G/L

7.3 Buffer

860