Printing a QC Setup Report

Use this procedure to print a QC setup report for any QC file.

1.

Access the QC File screen from the Menu screen:

a. Select

5 Operating Setup

and press

Enter

.

b. Select

1 QC Setup

and press

Enter

.

c. Select

1 QC Files

and press

Enter

.

2.

Select the file number that you want to print.

3.

Press

Print QC Setup

.

The QC Setup report prints. Figure 5-34 shows an example of a QC Setup
Report.

Figure 5Ć34.ā QC Setup Report

............. . . . . . . .

. . . . . . . . . . . . . . . . . . . .

.........................

840

850

QC SETUP REPORT APR 12 1994

11:42

SYSTEM 850-1001

GLOBAL QC SETUP

Auto ID is: On

Auto accept QC: Off

QC FILE SETUP

File 1

QC ID: 473842

Level: 2

Lot number: 41561

Expiration date: Dec 30 95

MEAN TARGET RANGE ACTION RANGE

pH 7.419 ( 7.399 - 7.439)( 7.389 - 7.449)

pCO

2

43.4 ( 38.4 - 48.4)( 35.9 - 50.9)

pO

2

103.1 ( 96.1 - 110.1)( 92.6 - 113.6)

Na+ 135.0 ( 130.0 - 140.0)( 127.5 - 142.5)

K+ 4.81 ( 4.31 - 5.31)( 4.06 - 5.56)

Ca++ 1.16 ( 1.06 - 1.26)( 1.01 - 1.31)

Cl- 102 ( 97 - 107)( 94.5 - 109.5)

Glucose 102.5 ( 97.5 - 107.5)( 87.5 - 117.5)

Lactate 0.92 ( 0.71 - 1.11)( 0.61 - 1.21)

ctHb 14.5 ( 13.5 - 15.5)( 13.0 - 16.0)

FO

2

Hb 1.2 ( 0.7 - 1.7)( 0.4 - 2.0)

FCOHb 95.5 ( 93.5 - 97.5)( 92.5 - 98.5)

FMetHb 0.4 ( 0.2 - 0.6)( 0.1 - 0.7)

FHHb 2.6 ( 2.1 - 3.1)( 1.8 - 3.4)

............................... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

860

These results
appear when an
845, 855, or 865
is interfaced.

..............

.............. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

CO-ox

4.

Press

Menu

.

5.

You can define another setup function or press

Exit Menu

to return to the

Ready screen.

1

1

Menu Code

5