Bayer HealthCare Rapidlab 800 User Manual

Page 142

Advertising
background image

2Ć44

Rapidlab 800 Operator's Manual

If you . . .

Then . . .

use the keypad or
keyboard

type the File Number and press

Enter

.

use the optional
bar code scanner

scan the bar code label on the QC ampule.

NOTE:

To ensure accurate sample identification, scan QC

ampule bar codes only after selecting Quality Control (QC)
from the Sample Type screen or when the QC File Information
form is displayed.

turn Auto ID on

system automatically determines which file and completes the
QC File Information screen.

NOTE:

If the pH or pCO

2

sensor or tHb is disabled or not in

calibration, Auto ID does not work.

do not complete
the screen

system stores results in QC File 14 and a message that contains
the sequence number and date and time of analysis is placed in
the status log. Refer to Editing QC Data, page 2-48, to move
the results to the correct file.

2.

Type the Operator ID and press

Done

.

If you want to change text in the QC File Information screen after you press
Done, press

Edit QC Data

. The QC File Information screen reappears. Make

your changes and press

Done

.

If you do not identify a QC file and you press Done, a message box appears
prompting you to identify the QC file. Press OK to return to the QC File
Information screen and complete the File Number field.

If you do not identify the QC file and press Done within 5 minutes, the QC sample
results are stored in the Discarded Data File. The system returns to the Ready
screen. Refer to Recalling QC Data, page 2-45, for instructions to recall, edit, and
print QC sample data after results are stored.

If you scan a bar code label that is not a QC ampule bar code, a message box
appears prompting you to scan a QC ampule bar code. Press OK to return to the
QC File Information screen and scan the correct QC ampule bar code label.

Procedural

Notes