Replacing the measurement module tubing – Bayer HealthCare Rapidlab 800 User Manual

Page 202

Advertising
background image

3Ć34

Rapidlab 800 Operator's Manual

Replacing the Measurement Module Tubing

Materials required:

S measurement module tubing

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

NOTE:

If your system has a CO-ox module, follow the procedure described for

the appropriate base model. For example, information identified for an 860 also
applies to an 865.

1.

Cut a piece of measurement module tubing to the appropriate length as shown
in Figure 3-19.

Figure 3Ć19.āăMeasurement Module Tubing Length

System

Tubing Length

840

15.2 cm (6 inches)

850

12 cm (4.75 inches)

860

11.43 cm (4.5 inches)

2.

Stop the system from the Menu screen:

a. Select

2

Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

3.

Push up the latches on the measurement module door and lift the door.

4.

Take the appropriate action.

If you have an . . .

Then continue with . . .

840

step 5.

850 or 860

step 6.

2

Menu Code

7