800 Series Operator's Manual

Figure HĆ4.āăInstalling the Gas Tank Regulator

Needle Valve

Adjustment

Knob

Gas Tank Seal

Regulator Nipple

Yoke

Screw

Gas Line

Yoke

Dowel Pins

Valve Stem

Valve Outlet

Dowel Pin Holes

Gas Tank

f. Attach the gas regulators to the gas tanks with the regulator nipple

engaging the opening in the tank valve.

g. Attach and tighten the yoke screws firmly.

h. Attach and tighten the tubing adapter fittings into each needle valve. Ensure

that the needle valve is fully open.