800 Series Operator's Manual

CAUTION:ĄĂ

Prolonged exposure to the 10% bleach solution damages the

reference sensor membrane. You must replace the reference sensor with a
test/blank ref sensor (TB5) while you complete the cleaning procedure. Do not
substitute a new reference sensor.

CAUTION:ĄĂ

Do not remove or return the sensors to the measurement module

without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.

3.

Replace the reference sensor with a test/blank ref sensor (TB5):

a. Remove the reference sensor from the measurement module and set it

aside.

b. Install a test/blank ref sensor (TB5) in the measurement module.

4.

Access the Condition screen from the Menu screen:

a. Select

2 Maintenance

and press

Enter

.

b. Select

2 Condition

and press

Enter

.

Wear safety glasses, gloves, and a laboratory coat when

handling the reagents.

5.

To clean the sample pathway:

a. Insert an aspiration adapter into the sample port and immerse the aspiration

adapter in the 10% bleach solution.

b. Press

Analyze

.

c. Remove the aspiration adapter when prompted.

d. Wait 5 minutes for cleaning to finish.

NOTE:ĄĂ

Press

Cancel

if you want to stop cleaning.

e. Press

No

. The Menu screen appears.

6.

Perform two additional wash sequences.

CAUTION:ĄĂ

Do not remove or return the sensors to the measurement module

without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.

7.

Replace the reference sensor:

a. Remove the test/blank ref sensor (TB5) in the measurement module and set

it aside.

b. Install the reference sensor in the measurement module.

!

!

2

Menu Code

2

WARNING

!