Measurement test, Printing stored signals – Bayer HealthCare Rapidlab 800 User Manual

Page 356

Advertising
background image

4Ć70

800 Series Operator's Manual

4.

Perform the appropriate action.

If you see the . . .

Then . . .

Temperature control system on
message

a. Press

Stop Test

.

b. Look at the status area to see if the D38

code has disappeared.

c. Allow the system to reach operating

temperature.

Temperature control system off
message

the system cannot reset the temperature control
system.

5.

Press

Exit Test

.

Measurement Test

Use this procedure to test the measurement electronics. You can:

S print stored signals from the last successful two-point calibration
S test the sensor circuit without a sensor and with a test/blank sensor

Printing Stored Signals

1.

Access the Measurement Test from the Menu screen:

a. Select

3 Troubleshooting

and press

Enter

.

b. Select

3 Measurement

and press

Enter

.

c. Select

1 Measurement

and press

Enter

.

2.

Press

Print Last Cal Mv

.

3.

Compare the results on the report to the expected values in the following
tables.

Sensor

Cal Signal

Expected Value

End Time, (sec)

Expected Value

pH

194.0 to 406.0 mV

20 to 90

pCO

2

–216.0 to +106.0 mV

25 to 90

pO

2

–0.129 to +2.071 nA

5 to 90

Na

+

29.0 to 131.0 mV

15 to 90

(Continued)

1

3

Menu Code