Analyzing capillary tube samples – Bayer HealthCare Rapidlab 800 User Manual

Page 106

Advertising
background image

2Ć8

Rapidlab 800 Operator's Manual

Analyzing Capillary Tube Samples

The following table lists the minimum volumes required by each system to analyze
capillary tube samples:

Table 2Ć2.āăMinimum Sample Volumes

System

Minimum Sample Volume

840
844, 845

0

90

µ

L

140

µ

L

850
854, 855

110

µ

L

160

µ

L

860
864, 865

125

µ

L

175

µ

L

NOTE:

If an 844, 845, 854, 855, 864, or 865 detects insufficient sample in the

measurement module, the system analyzes the sample as a microsample. If you
want the system to analyze a different set of parameters for the sample, press
Change Parameters to select another panel. Refer to Selecting Parameter Panels,
page 2-29, for more information.

BIOHAZARD:

Refer to Appendix A, Protecting Yourself from Biohazards, for

recommended precautions when working with biohazardous materials.

CAUTION:

Insert the fire-polished end of the capillary tube into the sample port to

prevent damage to the capillary seal.

1.

Prepare the sample and insert the fire-polished end of the capillary tube into
the sample port, as shown in Figure 2-3.

Figure 2Ć3.ā Inserting a Capillary Tube

Sample Port

Capillary Sample

!