4Ć48

800 Series Operator's Manual

D76 COĆox Electronics Error:

Problem

The system detects an error in the electronics system.

Qualifiers:

1 2

3

4

5

6

7

8

9

10 11 12

Solutions

If the qualifier is . . .

Then . . .

1, 2, 3, 5, 8, 11

perform the CO-ox Lamp On/Off test, and then perform a
one-point calibration.

Shut the system down and restart:

1.

From the Menu screen, select

7 System Utilities

and

press

Enter

.

2.

Select

3 Shutdown

and press

Enter

.

3.

Press

Yes

.

4.

Wait at least 1 minute and then disconnect the power cord
from the power source.

5.

Wait at least 10 seconds and then connect the power cord
to the power source.

4, 7, 9, 12

shut the system down and restart.

6

check the temperature control as described in
Temperature/pAtm Test, page 4-69.

Shut the system down and restart.

10

perform the Pump Functions Test as described in Pump
Functions Test
, page 4-58.

If the D code reappears, contact your Service Representative.

D76

7

Menu Code

3