Defining calibration gas values – Bayer HealthCare Rapidlab 800 User Manual

Page 468

Advertising
background image

800 Series Operator's Manual

5Ć58

1.

Access the Calibration Drift Limits screen from the Menu screen:

a. Select

5 Operating Setup

and press

Enter

.

b. Select

6 Calibration Setup

and press

Enter

.

c. Select

1

Drift Limits

and press

Enter

.

The Calibration Drift Limits screen appears, as shown in Figure 5-35.

Figure 5Ć35.ā Calibration Drift Limits Screen for an 850

Press Clear Entry to

delete text in a field.

2.

Move to the field that you want to change.

NOTE:ĄĂ

The number you enter establishes a range above and below the

expected value. For example, if you enter 2 as the drift limit, the range is plus
or minus 2 from the expected value.

3.

Type the new drift limit and press

Enter

.

4.

Repeat steps 2 and 3 for each parameter for which you want to enter or change
the drift limit.

5.

Press

Done

when you finish.

6.

You can define another setup function or press

Exit Menu

to return to the

Ready screen.

If you type an invalid value in a field and press Done, the Invalid Entry message
box appears. Press

OK

and type a valid entry in the field.

Defining Calibration Gas Values

Use this procedure to define the calibration gas values for the gases used during
calibration. Table 5-23 lists slope and cal ranges for

p

CO

2

and pO

2

.

1

6

Menu Code

5

Procedural

Notes