Shutting down the system, Preparing the 800 base model – Bayer HealthCare Rapidlab 800 User Manual

Page 607

Advertising
background image

Installation

HĆ15

8.

Carefully remove the foam end caps from the module.

9.

Remove the plastic bag surrounding the module.

10. Remove any tape used to secure parts during shipment.

Shutting Down the System

If you are installing the CO-ox module to an operational 800 series system,
perform the following procedure to shut down the system.

1.

Shut down the system from the Menu screen:

a. Select

7 System Utilities

and press

Enter

.

b. Select

3 Shutdown

and press

Enter

.

The Shutdown screen appears.

2.

Press

Yes

.

A message appears on the screen and on the roll printer, directing you to wait
before you disconnect the power.

CAUTION:ĄĂ

Do not unplug the system until the following message appears on

the screen:

...synching disks... done

This is an operating system message indicating that the system has
successfully completed the shutdown procedures. Unplugging the system
before this message appears can damage the system.

3.

When you see the operating system message, disconnect the power cord from
the power supply.

Preparing the 800 Base Model

This procedure refers to right and left as you look at the rear of system.

1.

Disconnect the power cord from the 800 base model.

NOTE:ĄĂ

Note the port location of each cable to ensure cables are reconnected

correctly.

2.

Disconnect the cables from the ports at the rear of the 800 base model. Refer
to Figure H-5.