Stopping the system – Bayer HealthCare Rapidlab 800 User Manual

Page 233

Advertising
background image

3Ć63

Maintaining the System

Stopping the System

Use this procedure to stop the system when you perform maintenance activities
such as replacing components. Stopping the system discontinues all fluidic
activities such as calibrations and sample analyses.

CAUTION:

Because no fluids reach the sensors while the system is stopped,

stopping the system for a prolonged period of time may affect the performance of
the sensors.

1.

Stop the system from the Menu screen:

a. Select

2 Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

2.

Perform the required maintenance task.

3.

Press

Continue

.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

4.

Take the appropriate action.

If . . .

Then . . .

you want to perform a
two-point calibration

press

Yes

.

you do not want to initiate a
two-point calibration

a. Press

No

.

b. Press

Home

to return to the Ready screen.

If calibrations were scheduled while the system was stopped, the system performs
the calibrations when you exit the stopped mode.

!

2

Menu Code

7

Procedural

Notes