Rapidlab 800 Operator's Manual

l. Disconnect the sample tubing from the CO-ox pump.

m. Connect one end of the replacement tubing to the outlet on the sample

chamber and the other end to the CO-ox pump.

3.

Replace the waste tubing:

a. Disconnect the waste tubing from the CO-ox pump.

b. Disconnect the waste tubing from the inlet on the reagent manifold on the

base model.

c. Gently pull the waste tubing through the hole in the FD5 housing.

d. Thread the replacement tubing through the hole in the FD5 housing.

e. Connect the replacement tubing to the inlet on the reagent manifold.

f. Connect the other end of the replacement tubing to the CO-ox pump.

4.

Press

Continue

.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

5.

Press

No

.

6.

Perform a one-point calibration from the Menu screen:

a. Select

1 Calibration

and press

Enter

.

b. Select

1 One-point

and press

Enter

.

1

Menu Code

1

(from the Main Menu)