HĆ24

800 Series Operator's Manual

Figure HĆ6.āăRemoving the Sample Port

Drip Tray

Sample Door

Retainer Ring

Mount

Sample Port

O-ring

4.

Grasp the sample port and drip tray and pull it to the right to remove it.

NOTE:ĄĂ

The sample port and the drip tray are one piece.

Wear safety glasses, gloves, and a laboratory coat when

handling the reagents.

5.

Clean any deposits on the sample port, the drip tray, and the mount with a
lint-free swab moistened with a 10% solution of household bleach.

6.

Rinse the sample port, the drip tray, and the mount with reagent water.

NOTE:ĄĂ

Ensure that the three O-rings are in place.

7.

Reinstall the sample port, matching the tab on the sample port to the notch in
the retainer ring.

8.

Push the tab on the retainer ring away from you until it locks in place.

9.

Reinstall the sample door, ensuring that it snaps in place.

10. Press

Continue

.

11. Press

No

.

12. Press

Home

to return to the Ready screen.

Cleaning the Exterior Surfaces

Wear safety glasses, gloves, and a laboratory coat when handling

the reagents.

1.

Using a 10% solution of household bleach, wipe all exterior surfaces
including:
S the sample entry components and the drip tray
S the waste area

WARNING

WARNING