Page 192

Advertising
background image

3Ć24

Rapidlab 800 Operator's Manual

Figure 3Ć10.āăRotating the Sample Tubing

Bracket

Sample Probe

Sample Tubing

Probe Mount

Sample Tubing Connector

5.

Disconnect the tubing from the sample probe as shown in Figure 3-11.

Figure 3Ć11.āăRemoving the Sample Tubing

Sample Probe

Sample Tubing

Probe Mount

Sample Tubing Connector

6.

Discard the tubing according to your laboratory’s biohazard protocol.

7.

Orient the new sample tubing as shown in the diagram on the system.

8.

Connect the new sample tubing to the sample probe.

9.

Place the sample tubing in the bracket on the probe mount.

10. Rotate the sample probe toward the system.