800 Series Operator's Manual

5Ć32

2.

Change the date and time.

If you want to

change the . . .

Then . . .

Date

a. Move the cursor to the Date field.

b. Type the month, date, and year in the selected format. Press

Enter

after you complete the field. The cursor moves to the

Time field.

Time

a. Move the cursor to the Time field.

b. Type the hour and minutes in the format, HH:MM. You can

enter 0 – 23 hours and 0 – 59 minutes.

c. Press

Enter

after you complete the field.

Date Format

a. Move the cursor to the required format.

b. Press

Enter

.

The date is reformatted, if necessary.

3.

Press

Done

when you finish.

If you changed the date or format, the Updating Database message box
appears.

4.

Define another setup function or press

Exit Menu

to return to the Ready

screen.

If you press Done before completing all required fields, the Incomplete Data Entry
message box appears. Press

OK

and complete the required fields.

If you type an invalid value in a field and press Done, the Invalid Entry message
box appears. Press

OK

and type a valid entry in the field.

To type a single number in a date or time field, you must precede the number with
a zero. For example, to enter June as the date, you must enter 06 in the month field.

Selecting Languages

Use this procedure to select the language that the system uses to present
information on the screen and in printed reports.

1.

Access the Service Setup screen from the Menu screen:

a. Select

8 Service Setup

and press

Enter

.

b. Select

3 Language Selection

and press

Enter

.

2.

Select the appropriate language and press

Enter

.

3.

Press

Done

.

4.

You can define another setup function or press

Exit Menu

to return to the

Ready screen.