D19 fluid detector error – Bayer HealthCare Rapidlab 800 User Manual

Page 314

Advertising
background image

4Ć30

800 Series Operator's Manual

D19 Fluid Detector Error:

Problem

Fluid detector millivolt (mV) reading is beyond predefined limits.

Qualifiers:

FD1

FD1A

FD2

Solutions

Check the sample path in the measurement module for obstructions or leaks.
Ensure that O-rings are in place on the sensors and that the sensors are aligned
correctly as described in Removing and Checking the Sensors on page 4-65.

If you see obstructions, remove them as described in Removing Obstructions
from the Measurement Module
, page 4-82.

Perform the Fluid Detector Test as described in Fluid Detector Test, page 4-66.

Deproteinize the sample path as described in Deproteinizing the Sample Path in
Section 3, and repeat the Fluid Detector Test.

If the D code reappears, contact your Service Representative.

D19