Cleaning the sample ground/temperature sensor – Bayer HealthCare Rapidlab 800 User Manual

Page 219

Advertising
background image

3Ć49

Maintaining the System

Cleaning the Sample Ground/Temperature Sensor

Materials required:

S 0.022 inch diameter clot-removal line (to fit through a 0.030 inch ID pathway)
1.

Stop the system from the Menu screen:

a. Select

2

Maintenance

and press

Enter

.

b. Select

7 Stop System

and press

Enter

.

2.

Push up the latches on the measurement module door and lift the door.

3.

Push the spring-loaded latch to the right as shown in Figure 3-31.

CAUTION:

Do not remove or return the sensors to the measurement module

without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.

4.

Grasp the tab on the sample ground/temperature sensor and pull the sensor up
and out of the measurement module.

Figure 3Ć31.āăRemoving the Sample Ground/Temperature Sensor from an 850

Push the springĆloaded

latch to the right.

Grasp the tab on the sample ground

sensor and pull it up and out of the

measurement module.

5.

Push a clot removal line through the sensor to clear away any obstructions as
shown in Figure 3-32.

Figure 3Ć32.āăCleaning the Sample Ground/Temperature Sensor

Clot Removal Line

OĆring

2

Menu Code

7

!