Bayer HealthCare Rapidlab 800 User Manual

Page 278

Advertising
background image

3Ć106

Rapidlab 800 Operator's Manual

3.

Disconnect the power cord from the side panel of the system.

4.

Remove the air filter cover.

5.

With a small, flat-blade screwdriver, gently pry open the fuse compartment
door at the top, as shown in Figure 3-72.

6.

Remove the fuse holders from the compartment.

Figure 3Ć72ā.ăReplacing the System Fuses

Power Input

Fuse Compartment Door

Air Filter Cover

Fuse

Fuse Compartment

Fuse Holder

Voltage Bobbin

Fuse compartment

door in open position.

Fuse Compartment Door Notch

Voltage Selection Bobbin

Window

7.

Remove the old fuses and install new fuses with the correct rating.

Refer to Table 3-2 to identify the correct fuse.

8.

Slide the fuse holders into the fuse compartment with the arrows pointing to
your right.

9.

Close the fuse compartment.

10. Ensure that the voltage selection bobbin is in place and the correct setting is

visible through the window.

11. Reinstall the air filter cover.